Eutech Instruments Turbidimeter TN-100 Waterproof Portable Meter
|
|
Bookmark Eutech Instruments Turbidimeter TN-100 Waterproof Portable Meter |
About Eutech Instruments Turbidimeter TN-100 Waterproof Portable MeterHere you can find all about Eutech Instruments Turbidimeter TN-100 Waterproof Portable Meter like manual and other informations. For example: review.
Eutech Instruments Turbidimeter TN-100 Waterproof Portable Meter manual (user guide) is ready to download for free.
On the bottom of page users can write a review. If you own a Eutech Instruments Turbidimeter TN-100 Waterproof Portable Meter please write about it to help other people. [ Report abuse or wrong photo | Share your Eutech Instruments Turbidimeter TN-100 Waterproof Portable Meter photo ]
Manual
Download
(English)
|
Eutech Instruments Turbidimeter TN-100 Waterproof Portable Meter
User reviews and opinions
| dbrodbeck |
3:03am on Saturday, October 23rd, 2010 ![]() |
| If you what a cycle computer this one is the bizz gives you great training data then put it on google earth and see your ride with the gps a must have... | |
| allpurpose |
5:05am on Tuesday, October 19th, 2010 ![]() |
| Altho this is the most expensive Garmin GPS system for the bike, I feel it has been worth the money from day one. I use the product to get the general GPS stats and derived things such as speeed and distance. However. | |
| texbmex |
6:35am on Sunday, October 17th, 2010 ![]() |
| Overall**** The Garmin Edge 705 is more than a bike computer. It is a full satellite navigation system for a bicycle. Available in 705. Assisting *It seems Garmin has settled down on its well designed training model, and no further development even in Edge 705. | |
| graupner |
8:07am on Saturday, September 4th, 2010 ![]() |
| The Garmin Edge is unique in design and handling. Self explanatory. Great support from Garmin and a lot of FOC gimmicks available on the internet. Does almost everything for fitness monitoring and with a built in GPS so you always know where you are Very compact design and an easy to read screen ... | |
| Leumis |
1:03am on Thursday, August 5th, 2010 ![]() |
| Paid a premium for this GPS compared to comparable models. Easy to use and works great Lack of flexibility good quality. quality support | |
| Pashmen |
4:46am on Sunday, June 20th, 2010 ![]() |
| Ability to see your ride afterwards in a number of different ways. It takes ages for the GPS to find Satellites. The base map is rubbish. The Garmin Edge is unique in design and handling. Self explanatory. Great support from Garmin and a lot of FOC gimmicks available on the internet. | |
| flo1979ro |
1:31am on Wednesday, June 9th, 2010 ![]() |
| It worth but with so restrictions I have purchase a second Egde 705 unit, not because I really love this model, but I cannot live without my gadgets. | |
| phaidros |
2:46am on Thursday, May 13th, 2010 ![]() |
| Garmin Edge 705 GPS-Enabled Cycling Computer Extremely happy with the Garmin Edge 705. It was relatively easy to install. Edgy I like it a lot. The only thing I would like to change is the maps. I use the Edge for biking. | |
| Gailor |
3:17pm on Saturday, March 27th, 2010 ![]() |
| Great for MTB This product is great fir training. Once you setup a gamin connect account you can go online and see how you did. Great device | |
Comments posted on www.ps2netdrivers.net are solely the views and opinions of the people posting them and do not necessarily reflect the views or opinions of us.
Documents
Instruction Manual
TN-100/T-100
Portable Turbidimeter
Technology Made Easy.
68X357701 Rev. 6 Mar 11
Part of Thermo Fisher Scientific
Preface
This manual functions in two ways: first as a step by step guide to help you operate the waterproof TN-100/ T-100 turbidimeter; second, it serves as a handy reference guide. It contains basic instructions that you must follow during the operation, care and maintenance of the instrument. The safety protection provided by this equipment may be impaired if it is used in a manner not described in this manual. It is recommended that all operators should read this manual prior to working with this instrument. Eutech Instruments/ Oakton Instruments cannot accept any responsibility for damage or malfunction to the meter caused by improper use of the instrument. The information presented in this manual is subject to change without notice as improvements are made, and does not represent a commitment on the part of Eutech Instruments Pte Ltd/ Oakton Instruments. Note: Eutech Instruments Pte Ltd/ Oakton Instruments reserves the right to make improvements in design, construction, and appearance of products without notice.
Copyright 2006 All rights reserved. Eutech Instruments Pte Ltd Oakton Instruments
Table of Contents
1. Overview.... 1
1.1 1.2 1.3 1.4 Unpacking TN-100/ T-100 Meter and Accessories.. 1 Display.... 2 Keys and Functions.... 3 Battery Installation.... 4
2. Turbidity Calibration... 5
2.1 2.2 2.3 2.4 Calibration Standards.... 5 Indexing..... 5 Calibration Procedure... 6 Restoring Factory Calibration... 8
3. Turbidity Measurement... 10
3.1 3.2 3.3 General Information.... 10 Turbidity Measurement Procedure... 10 Single-Shot or Continuous Measurement... 13
4. Troubleshooting Guide... 15 5. Routine Maintenance.. 16
5.1 Vials Handling, Cleaning and Care... 16
6. Accessories.... 17 7. Specifications... 18 8. Addendum 1: Turbidity... 19
8.1 8.2 8.3 8.4 8.5 Definition.... 19 Why Is It Important?.... 19 Measurement Principle.... 19 Nephelometric Turbidity Units (NTU)... 20 Indexing a Vial.... 20
9. Addendum 2: Guide to Good Measurement Technique. 21
9.1 9.2 9.3 9.4 9.5 9.6 9.7 Maintain sample vials in good condition... 21 Match Sample Vials.... 21 Degassing.... 21 Timeliness of Sample... 22 Other Important Sampling Techniques... 22 Calibration.... 22 Dilution.... 23
10. 11.
Warranty... 24 Return of Items... 24
TN-100/ T-100
1. Overview
Thank you for selecting the waterproof portable TN-100/ T-100 turbidimeter. The TN100/ T-100 allows you to measure turbidity of an aqueous sample in the field. This instrument operates on the nephelometric principle of turbidity measurement and is designed to meet the criteria specified in ISO 7027 and DIN 27027 standards (see Measurement Principle on page 19).
1.1 Unpacking TN-100/ T-100 Meter and Accessories
The table below indicates the items that you should find in your turbidimeter shipment.
Item 1. Portable TN-100/ T-100 Turbidimeter with 4 AAA batteries 2. Instruction Manual 3. Instrument Carrying Case 4. Light Shield Cap 5. Calibration Set (0.02, 20.0,100, 800 NTU Standards) 6. Empty Vials 7. Plastic bottle (empty for collecting sample) 8. Silicone Oil 9. Lint free cloth Quantity 1
Remove TN-100/ T-100 turbidimeter from the packing carton. Carefully inspect all items to ensure that no visible damage has occurred during shipment. If the items you received do not match your order, please contact your nearest distributor immediately. WARNING: Extra care should be taken when unpacking, opening, and handling the calibration standards and sample vials. Surface scratches or finger smudges on the vial surface may cause measurement errors. Handle these items by their caps only. Batteries provided with the meter package are to be installed prior to use. See Section 1.4 - Battery Installation on page 4. Figure 1 depicts the meter. The three main components of the instrument are the sample well, the display, and the keypad. The following sections describe the functionality of the display and the keypad. The proper use of the instrument and the sample well are discussed in later sections.
Sample Vial Sample Well Index Mark
Liquid Crystal Display
Keypad
Figure 1: Parts of TN-100/T-100 Turbidimeter
1.2 Display
All the LCD segments and annunciators that can appear on the display are shown in Figure 2. The display is used for reporting the turbidity reading and to provide guidance for the operation of the instrument. In addition, the display has several other annunciators that are used to communicate error messages and provide user guidance.
Calibration Standards Battery Indicator
CAL 13 CAL 24
mg/l NTU ppm %pH
Units of Measurement
14-segment Liquid Crystal Display
Figure 2: Customized LCD with all annunciators
1.3 Keys and Functions
The keypad has five keys: ON/OFF, CAL, , , and READ/ENTER.
ON OFF
Functions
Powers on and shuts off the meter. With the auto switch-off feature the meter automatically shuts off 20 minutes after last key press.
Initiates the calibration mode of the meter. When pressed, the meter is set to accept the first calibration standard. It is also used to exit the calibration mode if the user does not want to follow the complete calibration procedure. When in measurement mode, the READ/ENTER key is used to perform a measurement. Single-shot measurements: When the key is pressed and released immediately (a quick key stroke of less than 0.3 seconds), the display will blink [-Rd-] for 10 times and then display the measured value. Continuous measurements: If the READ/ENTER key is pressed and held, the instrument will perform a continuous measurement during which the display is updated every 2 seconds. This can be used for indexing vials. (This function is not available in calibration mode). After the READ/ENTER key is released, the instrument will automatically perform a single-shot measurement. Active only during calibration mode. These keys are used to select the calibration point in an incremental/ decremental manner. NOTE: This meter is fitted with an auto incrementing feature, i.e. after the successful calibration of one point it auto selects the next calibration point, or automatically exits the calibration mode after the fourth calibration point.
READ ENTER
WATERPROOF
Figure 3: Keypad
1.4 Battery Installation
Four AAA-sized batteries are included in the packaging: 1. Use a Philips screwdriver to remove the two screws holding the battery cover. See Figure 4. 2. Remove the battery cover. 3. Insert the batteries. Follow the diagram inside the cover for correct polarity. 4. Replace the battery cover onto its original position using the two screws removed earlier. 5. The meter is now ready to operate.
Philips Screws
Battery cover
Four 'AAA'-sized batteries
Figure 4: Battery Installation NOTE: Dispose used batteries in accordance with your local regulations.
2. Turbidity Calibration
The TN-100/ T-100 turbidimeter was calibrated and tested prior to leaving the factory. Therefore, it is possible to use the instrument directly out of the box. However, recalibration of the instrument is recommended to help you become familiar with the operation of the instrument and the calibration procedures. In addition, re-calibration is recommended at least once every month for optimum accuracy. It is recommended that you perform a full calibration using all 4 standards provided to ensure full-range accuracy. However, the TN-100/ T-100 turbidimeter provides flexibility for you to calibrate at selected ranges suitable for your application. In addition, the instrument comes with a light shield cap to shield off stray light during calibration and measurements.
2.1 Calibration Standards
We recommend that you use the following materials during calibration to achieve the accuracy stated in this manual: CAL 1: 800 NTU Calibration Standard CAL 2: 100 NTU Calibration Standard CAL 3: 20.0 NTU Calibration Standard CAL 4: 0.02 NTU Calibration Standard
It is well known that diluted Formazin is unstable. If you choose to use Formazin to calibrate the instrument, ensure that you are using a fresh stock suspension of Formazin to achieve the accuracy quoted for the instrument. Calibration standards offered are more stable than Formazin. For expiry date of calibration standard, please refer to the expiration date indicated on the cap label.
It is important that the calibration standards are not violently shaken or agitated because air entrapment in the fluid introduces an error factor during calibration which subsequently will lead to an inaccurate measurement. Also, do not store in freezing temperatures which causes irreversible shrinkage of the standards particles thus resulting to inaccurate calibration and measurement.
2.2 Indexing
Due to the high quality of the glass vials provided, indexing is not required. You only need to align the mark on the vial with the mark on the meter. However, in order to achieve a better accuracy of the measurement, you can proceed with indexing of the vials. See Section 8.5 - Indexing a Vial on page 20 for more information.
2.3 Calibration Procedure
1. Place TN-100/ T-100 turbidimeter on a flat and level surface. 2. Insert the CAL 1 standard (800 NTU) into the sample well, aligning the mark on the vial with the mark on the meter. See Figure 10 on page 12. 3. Press down vial until it snaps fully into the instrument. 4. Cover the vial using the light shield cap. 5. Press ON/OFF key to switch the on the meter. The meter goes to measurement mode after the power-up sequence. 6. Select the calibration function of the instrument by pressing the CAL key once. The CAL annunciator blinks momentarily and the meter prompt for the first calibration standard CAL NTU. 7. Press READ/ENTER key. 8. The CAL NTU annunciator blinks for about 12 seconds while the instrument performs calibration of CAL 1 point. When the instrument has completed calibration for this point, it prompts you to insert the next calibration standard into the sample well CAL NTU. 9. Repeat the calibration for CAL 2, CAL 3 & CAL 4 calibration standards. 10. After you successfully calibrate the CAL 4 standard (0.02 NTU), the display shows STbY. 11. The meter is now ready for measurement. Figure 5 shows the complete calibration sequence.
NOTES: 1. If you wish to exit the calibration mode you may do so at the end of any step by pressing the CAL key. The meter accepts only the values calibrated prior to exiting. 2. You can skip a calibration point by pressing or keys and move on to the next calibration point. 3. After a successful calibration of one point, the meter automatically selects the next calibration point. The meter automatically exits calibration mode after the fourth calibration point. 4. If an error occurs during calibration, the display shows an error message. The meter aborts calibration and returns to the measurement mode without saving the last calibration value. 5. For a list of error messages, refer to Section 4: Troubleshooting Guide on page 15.
Measurement Mode
Flashes about 5 times
20.0 NTU
Place the CAL 3 standard into the sample well
Flashes about 10 times
800 NTU
Place the CAL 1 standard into the sample well
0.02 NTU
Place the CAL 4 standard into the sample well
100 NTU
Place the CAL 2 standard into the sample well
Figure 5: Calibration Sequence
2.4 Restoring Factory Calibration
The TN-100/ T-100 turbidimeter allows you to reset the meter back to the factory default calibration values. This feature is extremely useful when there are errors in calibration or when you have new calibration standards. 1. With the meter switched off, press and hold key. 2. Switch on the meter using ON/OFF. Release key when URST (User-Reset) appears in the display. 3. The display shows URST for about 2 seconds and then shows No. 4. Use or key to select between YES or No. YES = To restore meter back to factory calibration values No = To retain last calibrated values
Figure 8: Sample Vial
8. The sample vial is now ready to be inserted into the sample well of the meter for measurement.
Figure 9: Apply a thin film of silicon oil
Measurement Procedure
1. Place TN-100/ T-100 turbidimeter on a flat and level surface. 2. Place the sample vial inside the sample well and align the vials index mark with the meters index mark. See Figure 10.
Light shield cap for covering the vial during measurements
Align index mark on the sample vial with the index mark on the TN-100/ T-100 turbidimeter and push till it is fully snapped in IMPORTANT: Place meter on a flat and level surface. DO NOT hold it on hands while operating the meter. It may lead to inaccurate readings
Figure 10: Align sample vials index mark with meters index mark 3. Push the vial until it is fully snapped in. 4. Cover the vial with the light shield cap. 5. Turn on the meter by pressing the ON/OFF key. 6. After the power-up sequence, the meter goes to measurement mode and the display blinks --Rd-- for about 10 times. See Figure 11. 7. The measured reading appears in the display. 8. If necessary, place the second sample vial into the sample well. Remember to align the vials mark with the meters index mark. 9. Press READ/ENTER key. The display blinks --Rd-- for several times and measured reading appears. 10. Repeat steps 1 through 9 for all of your samples.
SAMPLE
Unknown NTU
Place sample vial into the sample well
Figure 11: Reading Turbidity Value
NOTES: 1. NEVER pour liquid directly into the sample well of the instrument. Always use a vial. The instrument will only accurately measure the turbidity of a sample when vials sealed with the black caps are used. The black cap serves as both seal and a light shield. 2. NEVER attempt to clean the sample well. The optics may be damaged. 3. For battery conservation, the instrument automatically powers off 20 minutes after the last key pressed.
3.3 Single-Shot or Continuous Measurement
You can use TN-100/ T-100 turbidimeter to take a single reading or perform continuous measurement. The latter is only used for indexing the vials. See Section 8.5 - Indexing a Vial on page 20 for more information.
For Single-shot Measurement:
1. Make sure the meter is sitting on a flat and level surface and is in measurement mode. The display shows the last measured value or STbY after exiting calibration mode. 2. Place sample vial in the sample well. 3. Cover the vial with the light shield cap. 4. Press READ/ENTER key and release immediately (<0.3 seconds). See Figure 12. 5. The display blinks --Rd-- for about 10 times and then display the measured value.
Press key and release immediately
Figure 12: Single-shot Measurement
For Continuous Measurement:
1. Make sure the meter is sitting on a flat and level surface and is in measurement mode. The display shows the last measured value or STbY 2. Place the sample vial in the sample well. 3. Press READ/ENTER key and hold. See Figure 13 4. Wait for the reading to stabilize before rotating the sample vial. NOTE: During continuous measurement, the display is updated every 2 seconds. The displayed reading may not be the actual turbidity value. For accurate measurement, use single-shot measurement. You can rotate the sample vial for indexing purpose. See Section 8.5 - Indexing a Vial on page 20 for more information. 5. Once you release READ/ENTER key, the meter automatically performs a singleshot measurement.
Press key and hold
Reading flashes and updates every 2 seconds
Figure 13: Continuous Measurement NOTES: 1. After a measurement is completed and the display is updated, a 4second recovery time occurs before the meter can perform another function. If any key is pressed during the recovery time, the meter will perform the corresponding action at the end of the recovery period. 2. When performing a measurement, if the meter detects stray light exceeding the amount equivalent to 0.02 NTU in the low range, the measurement is immediately aborted and an error message [ERR 8] is displayed. Ensure the vial sits properly into the sample well. Press the ENTER/READ key to re-do the measurement.
4. Troubleshooting Guide
The TN-100/ T-100 turbidimeter routinely performs self-diagnostics, and will automatically generate messages to provide you with specific diagnostic information. These messages are for your use and do not indicate a reduction in the performance of the instrument or a failure of any component in the instrument, unless otherwise stated in this list. LCD Message
Description
Calibration Error. The meter is unable to recognize the 800 NTU Calibration Standard. Calibration Error. The meter is unable to recognize the 100 NTU Calibration Standard. Calibration Error. The meter is unable to recognize the 20.0 NTU Calibration Standard. Calibration Error. The meter is unable to recognize the 0.02 NTU Calibration Standard. Calibration Error. There is not sufficient signal to achieve the appropriate resolution in the 01000NTU range. General Calibration Failure. There is not sufficient signal to achieve the appropriate resolution in the 0100NTU range. General Calibration Failure. There is not sufficient signal to achieve the appropriate resolution in the 020 NTU range. Excessive stray light detected.
6. Accessories
The items shown below are recommended accessories and replacement parts for the instrument. Item Description
Waterproof TN-100/ T-100 portable turbidimeter with set of 4 calibration standards (800, 100, 20.0 & 0.02 NTU), set of 3 sample vials, lint-free cloth, silicone oil, batteries all in a rugged carrying case. Calibration Set for normal operation (includes 800, 100, 20.0 & 0.02 NTU Standards) Sample Vials pack of 3 vials Silicone Oil 10ml
Eutech Instruments Order Code
EC-TN100
Oakton Instruments Order Code
35635-00
ECTN100CALKT ECTN100CUVKT ECSILICONEOIL
35635-50 35653-55 ~
To order any accessory or replacement part, please contact the nearest distributor or Eutech Instruments/ Oakton Instruments.
7. Specifications
Parameter
Measurement Method Measurement Range Automatic Range Selection
Specification
ISO 7027 compliant nephelometric method (90) 0 to 1000 NTU 0.01 19.99 NTU 20.0 99.9 NTU NTU 0.01 NTU (0 19.99 NTU) 0.1 NTU (20 99.9 NTU) 1 NTU (NU) 2% of reading 1 LSD for 0 to 500 NTU; 3% of reading 1 LSD for 501 to 1000 NTU. 0.01 NTU or 1% of reading, whichever is greater with gel samples < 6 seconds for full step change 0.02 NTU; 20 NTU; 100 NTU; 800 NTU EPA-approved polymer-based primary standards Infrared-emitting diode (850 nm wavelength) > 1,000,000 tests Silicon photovoltaic < 0.02 NTU 4-digit 14-segments customized liquid crystal display with annunciators Borosilicate glass with screw caps, fill line and indexing mark. 51 (H) x 25 (Dia) mm ( 2 x 1 in) 10 ml (0.33 oz) 0C to 50C (32F to 122F) 0C to 50C (32F to 122F) 0-90% RH, non-condensing at 30C (86F) 4 x AAA Alkaline Batteries > 1200 readings ABS Plastic / IP67 rated Pollution Degree 2 Meter: 200 g (7 oz) Meter with case: 1.25 kg (2.75 lb) Meter: 6.8 (W) x 15.5 (L) x 4.6 (H) cm; (2.7 x 6.1 x 1.8 in) Meter with Case: 16 (W) x 35 (L) x 12 (H) cm; (6.3 x 13.8 x 4.7 in)
Resolution
Accuracy Repeatability Response Time Calibration Standards Standardization Light Source Light Source Life Detector Stray Light Display Sample Cells (Vials)
Sample Volume Required Operating Temperature Range Sample Temperature Range Operating Humidity Range Power Supply Battery Life Enclosure Type & Rating: Insulation Rating Weight: Dimensions
8. Addendum 1: Turbidity
8.1 Definition
Turbidity is defined as an expression of the optical property that causes light to be scattered and absorbed rather than transmitted in straight lines through the sample.1 That is, turbidity is the measure of relative sample clarity, not colour. Water with cloudy or opaque appearance will have high turbidity, while water that is clear or translucent will have low turbidity. High turbidity value is caused by particles such as slit, clay, microorganisms, and organic matter. By definition, turbidity is not a direct measure of these particles but rather a measure of how these particles scatter light.
8.2 Why Is It Important?
For drinking water application, a turbidity value may give an indication of presence of bacteria, pathogens, or particles that can shelter harmful organisms from disinfection process. Therefore, turbidity measurement is particularly useful for water treatment plants to ensure cleanliness. In industrial processes, turbidity can be part of quality control measure to ensure efficiency in treatment or manufacturing process.
8.3 Measurement Principle
There are two internationally accepted standard specifications for turbidity measurement. These are the international standard ISO 7027 and the US EPA method 180.1. Basically the ISO 7027 is a more stringent standard and requires the use of a monochromatic light source. It also governs the design of a turbidimeter in the following areas: (1) Light sources wavelength; (2) Light sources spectral bandwidth; (3) Measuring angle; (4) Aperture angle in water sample; (5) Distance traversed by incident light and scattered light within the sample; (6) Calibration standard. The TN-100/ T-100 turbidimeter follows the ISO 7027 standard whose specification allows for greater reproducibility of the measured values and greater agreement between other measuring instruments.
Provided by Standard Methods for the Examination of Water and Wastewater, APHA, AWWA and WPCF, 16th Edition, 1985. 19
Figure 14 shows the waterproof TN-100/ T-100 turbidimeter basic optical system. It includes a light source and a detector to monitor the light scattered at 90 with respect to the incident beam.
Scattered Light
Lamp Lens Glass Sample Vial
Transmitted Light
Figure 14: Basic Nephelometric arrangement for turbidity measurement
8.4 Nephelometric Turbidity Units (NTU)
Nephelometric Turbidity Units (NTU): Unit of measure used when relating the light scattered by a liquid media to the light scattered by a known concentration of a standard solution. This unit of measure is recognized as a measure of the optical clarity of an aqueous sample. NTU is the accepted unit of measurement for turbidity. Another unit commonly used to measure turbidity is Formazin Turbidity Unit (FTU). The two units of measure of turbidity are equivalent: 1 NTU = 1 FTU.
8.5 Indexing a Vial
The United States Environmental Protection Agency (US EPA) recommends that vials used for turbidimeter calibration or sample measurement be indexed. To index a sample vial, slowly rotate the vial throughout one complete revolution (360). While rotating the sample vial, observe the display and locate the position that the vial is in which provides the lowest turbidity reading. This position is the indexed position of the vial. Mark this position on the vial (not on the cap) against the mark on the meter. After indexing a vial, make sure the vial will always be placed inside the sample well in the indexed position.
9. Addendum 2: Guide to Good Measurement Technique
Turbidity is a very complex analytical measurement which can be affected by many factors. Some are inherent in the instruments design such as angle of detection, light beam aperture, incident beam wavelength and color sensitivity of the photocell. However, there are other factors such as stray light, air bubbles and care of vial, which can be prevented through proper care of equipment and accessories, and in the operating procedure for measurement. Here are some points you may want to note:
9.1 Maintain sample vials in good condition
Sample vials must be meticulously clean and free from significant scratches. It should be treated on the outside with a thin coat of silicone oil. This is to mask minor imperfections and scratches that may contribute to stray light. Sample vials should be handled only by the top to avoid dirt accumulation (or deposits) and fingerprints that might interfere with the light path. More information is found in Section 5.1 - Vials Handling, Cleaning and Care on page 16.
9.2 Match Sample Vials
Best accuracy and repeatability of turbidity measurement are achieved using a single, indexed vial. However, for more convenience, different vials can be used for measurement provided their readings with the same solution are matched. That is, the meter gives identical readings or within the specified repeatability and accuracy of the meter. Select a few vials. After the sample vials are cleaned, fill them with ultra-low turbidity water. Allow the sample vials to stand and for air bubbles to rise. Polish sample vials with silicone oil and take turbidity measurement at several points while rotating it in the sample well. Find the position where turbidity reading is the lowest and index it for each vial. Whenever these sample vials are used, use the indexed mark to position each vial into the sample well. Choose those vials that match the readings. NOTE: Not all vials can be matched because of some manufacturing variations.
9.3 Degassing
Air or other trapped gases should be removed before measurement. Degassing is recommended even if no bubbles are visible. There are three methods commonly used
Instruction Manual for degassing:
Addition of a surfactant: This involves adding a surfactant to the water samples to lower the surface tension of the water, thereby releasing trapped gasses. Application of a partial vacuum: Partial vacuum can be created by using simple syringe or vacuum pump. (This is only recommended for ultra-low turbidity measurement). Use of an ultrasonic bath: This may be effective in severe conditions or in viscous samples, but not recommended for ultra-low measurements.
Each of the method above has its own advantages and disadvantages. For instance, under certain sample conditions, the use of vacuum pump or ultrasonic bath may actually increase the presence of gas bubbles.
9.4 Timeliness of Sample
Samples should be measured immediately to prevent changes in particle characteristics due to temperature and settling. Temperature can affect particles by changing their behavior or creating new particles if precipitates are formed. Dilution water may dissolve particles or change their characteristics. It is recommended to take samples only when the turbidimeter is ready to be operated. Samples should not be drawn and allowed to sit while the instrument warms up or is being readied.
9.5 Other Important Sampling Techniques
1. Samples should not be violently shaken or agitated as particles can be broken apart or air may be entrapped into the fluid. Gentle agitation such as swirling the sample vial is advisable to reduce particle settling. 2. Sample vials should be used only with the instruments for which they were intended. Do not mix and match. 3. Perform a visual observation of the sample vial every time a measurement is made. Ensure that there are no visible bubbles in the sample and the vial is clean and free of scratches. 4. Samples entering the turbidimeter should be at the same temperature as the process flow samples. Changes in temperature can cause precipitation of soluble compounds and affect readings. 5. Sample vials should be evaluated with a low turbidity water (after cleaning) to determine if cells remain matched. If the evaluation determines that a cell is corrupted, discard the vial. It is recommended to conduct this evaluation weekly. 6. When in doubt or question about whether a sample vial is too scratched or stained, throw it away.
9.6 Calibration
1. Do not open the vials with calibration standards. 2. Check that the standards have not expired.
3. Make sure the calibration vials are free of dust, smudges and scratches before use. 4. Conduct the calibration in the same manner each time. Variations in how calibration is performed could yield inaccurate measurements. 5. It is very important that the user(s) who perform calibration have been trained to do so. Creating a Standard Operating Procedure (SOP) for the user(s) to read, learn, and practice may help to ensure accuracy.
9.7 Dilution
This dilution procedure is necessary only when your turbidity measurement is above 1000 NTU. 1. To measure the turbidity above 1000 NTU, dilute the sample with turbidity-free water. 2. Turbidity-free water can be obtained by filtering deionized water through a < 0.2m filter membrane with precision-sized pores. 3. Measure the volume of the sample before dilution and record the value in ml (Vs). 4. Take a known volume (Vd) of dilution water and add it to the sample. 5. Pour 10 ml of the diluted sample in a clean vial and measure the turbidity of the diluted sample. Record this value in NTU (Td). 6. Calculate the true turbidity (T) of the original sample - in NTU - using the following formula:
T = Td * (Vs + Vd) / Vs
Example: Dilute 20 ml of the original sample (whose turbidity is above 1000NTU) with 50 ml of dilution water. Measure the turbidity of the diluted sample. If the reading is 300 NTU, the turbidity of the original sample is 1050 NTU. (In this case: Td=300NTU, Vs=20ml, Vd=50ml, so T = 300 * (20+50) / 20 = 300*70/20 = 21000/20 = 1050)
10. Warranty
The TN-100/ T-100 meter is supplied with a 2-year warranty from manufacturing defects and calibration standards for 6 months. If repair or adjustment is necessary and has not been the result of abuse or misuse within the designated period, please return freight pre-paid and correction will be made without charge. Eutech Instruments/ Oakton Instruments will determine if the product problem is due to deviations or customer misuse. Out of warranty products will be repaired on a charged basis. Exclusions The warranty on your instrument shall not apply to defects resulting from: Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products
Waterproof Seal: Opening the instrument enclosure (excluding the battery compartment) may void the warranty.
11. Return of Items
Authorization must be obtained from our Customer Service Department or authorized distributor before returning items for any reason. A Return Goods Authorization (RGA) form is available through our Authorized Distributor. Please include data regarding the reason the items are to be returned. For your protection, items must be carefully packed to prevent damage in shipment and insured against possible damage or loss. Eutech Instruments/ Oakton Instruments will not be responsible for damage resulting from careless or insufficient packing. A restocking charge will be made on all unauthorized returns.
NOTE: Eutech Instruments Pte Ltd/ Oakton Instruments reserve the right to make improvements in design, construction, and appearance of products without notice.
For more information on Eutech Instruments/ Oakton Instruments products, contact your nearest distributor or visit our website listed below: Oakton Instruments P.O Box 5136, Vernon Hills, IL60061, USA Tel: (1) 888-462-5866 Fax: (1) 847-247-2984 E-mail: info@4oakton.com Web-site: www.4oakton.com Eutech Instruments Pte Ltd Blk 55, Ayer Rajah Crescent, #04-16/24 Singapore 139949 Tel: (65) Fax: (65) E-mail: eutech@thermofisher.com Web-site: www.eutechinst.com Distributed by:
Technology Made Easy.
EUTECH INSTRUMENTS PTE LTD Blk 55, Ayer Rajah Crescent #04-16/24, Singapore 139949 Republic of Singapore Tel: (65) 6770 2804; Fax (65) 6773 0836
Press Release
For Immediate Release Friday, 1-July-2003 Contact: William Soh, Marketing Manager (williams@eutechinst.com) Tel: +65-6770 2804; Fax: +65-6773 0836
TN 100 Waterproof Portable Turbidimeter
The Waterproof TN 100 from Eutech Instruments, is a compact and light-weight (200 g) waterproof hand-held turbidimeter designed for portability and durability. It uses nephelometric method with Infra Red LED which is compliant to ISO 7027/DIN EN 27027 requirements. The microprocessor-based turbidimeter features push-button operation, large easy-to-read display, and user-friendly message codes. It can even detect stray light! The TN 100 automatically recognizes calibration standards and even allows user to calibrate at selected points. Auto-ranging function senses the turbidity level and automatically adjusts to the appropriate measurement and accuracy. Measurement range is 0 to 1000 NTU with accuracy of 2% (0 to 500 NTU) or 3% (501 to 1000NTU). Resolution is 0.01 on the lowest scale. A rugged portable carrying case and waterproof housing ensure durability and ability to measure everywhere. The TN 100 includes everything needed for the measurement from batteries to 4 calibration standards (0.02, 20.0, 100 and 800 NTU), sampling bottle, lint-free cloth, silicone oil and 3 empty vials. With laboratory repeatability and accuracy now in a small palm-sized meter, the TN 100 is a valuable tool for field research and quality control. Eutech Instruments, an ISO 9001 certified company, is a major manufacturer in electrochemical instruments and markets its products world-wide. For more information, contact our distributors or visit website, www.eutechinst.com.
PRESS RELEASE
PAGE 1 / 1
EUTECH INSTRUMENTS
Tags
GT-S7350 AD-F810 23-32-50 KW Sport Plus Blackberry 8120 FS-V21RP RQ-SW20 RAV120 DSC-5 MS-2387AR Graphire CB-250 Air WAR TH-37PX80B PCG-FX503 CDX-GT350 CRW8824S OT-807 Gear Acid ICN 330 EM-120 793MG P4000 A75-S229 32LC42 Sinus 700 2043BWX Virtualdj 4 Steam Guide Mf4350D XA-950 PX3000 32PW6518 KX-TCD340 550 RA ERD24001W8 MF-JM53s8K DC211 EX-P600 DJM-1000 HM321HI DIN-1520 LS-K1863CL Monitor Plug-IN Plus 710 USB Nuvi 270 EMP-820 IC-F111 CDX-M770 C1 125 Encoder Serie E Easyshare M863 DDX5024 SCD 23 T135-2006 M181GN MG5050 Class 270 XTA C2 RTD170 ECR 6700 DX609I Perfection 4490 XR-CA630X SD770 IS Samsung 245B Tool XD 7 IN1 CS-CE9hkew Zigity Review C210-220 MOC ML2851ND-XAZ KX-TG1070SP FP-7718 Laserfax 920 ALL-IN-ONE Neobio 30 P3470 AG-DVC7 Gloss Gigaset S675 Alcatel-lucent 4068 FW-C85-21M Standard TX-NR5007 CFX-9850GC Plus 400DXN Photo R320 C18AHR 5761A Fujifilm Z30 BR-1600 CD Mouse G4 15 Express KX-FPC91 EWF1010
manuel d'instructions, Guide de l'utilisateur | Manual de instrucciones, Instrucciones de uso | Bedienungsanleitung, Bedienungsanleitung | Manual de Instruções, guia do usuário | инструкция | návod na použitie, Užívateľská príručka, návod k použití | bruksanvisningen | instrukcja, podręcznik użytkownika | kullanım kılavuzu, Kullanım | kézikönyv, használati útmutató | manuale di istruzioni, istruzioni d'uso | handleiding, gebruikershandleiding
Sitemap
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93 94 95 96 97 98 99 100 101